Optical
ACD Measurement with the Zeiss IOLMaster
1.
What
should I bear in mind for taking the measurements?
-
Focus onto focus point (smallest dot)
-
Make sure the dot is within the green square.
-
Avoid too strong extraneous reflections and room light.
-
The image of the lens should be well visible on the screen.
-
Front surface of the lens should be visible inside the pupil
2. Why
should the ACD be measured?
Most
formulas use axial length and corneal power for IOL-calculation only and
calculate a predicted ACD.
This
may cause problems in case of short or long eyes.
According
to J. Holladay 10% of short eyes have long ACD and 10% of long eyes short ACDs.
Therefore,
especially in these cases, the use of a biometric formula with measured ACD
should be better.
3. What is the principle of anterior
chamber depth measurement?
The measuring principle of anterior
chamber depth measurement is based on the "optical section" through
the anterior chamber by means of a slit lamp with subsequent image analysis.
The right eye is illuminated from the right; the left eye is illuminated from
the left at an angle of approx. 30° relative to the optical axis.
The measurement is performed by
image analysis.
The instrument measures the anterior-chamber depth (ACD) as it is usually
measured in biometry. Anatomically, this is the anterior chamber depth plus
corneal thickness.
4. What should I think of, if in
successive anterior chamber depth measurements large deviations in the range of
0.1 mm occur?
Naturally, patient accommodation
affects anterior chamber depth measurement and lens thickness. The patient
should be told to look relaxed at a distant point. To eliminate accommodation
and make measurements with small anterior chamber and small pupils easier, it
is possible to measure under a mydriatic.
5. Is
it possible to measure pseudophakic eyes with the anterior-chamber depth meter?
The measurement requires
scattered light from the lens. The IOL of freshly operated patients is hardly
scattering light. That's why measurements are not possible in this case.
6. How often may I measure an eye
per day using the anterior chamber depth-measuring device?
The number of measurements is not
limited. The light load on the patient remains uncritical even with a great
number of measurements. A few patients however find the white light/flickering
light too glaring.
7. What is the reproducibility on
the eye?
Reproducibility is better
than ±50 µm, this corresponds to a mean postoperative error
of less than 0.05 D.
In clinical use, mean
standard deviations of 31,9 µm (15 measurements, Dr.Haigis, University Eye
Hospital Wuerzburg/Germany) and 33,4 µm (20 measurements, Dr. Vogel, University
Eye Clinic, Mainz/Germany) were obtained. The variabilities between different
examiners were 37.7 µm (4 examiners, Dr.Haigis, University Eye Hospital
Wuerzburg/Germany) and 29.8 µm (5 examiners, Dr.Vogel, University Eye Clinic,
Mainz/Germany).
Safety criterion:
If the results of
the individual measurements differ by more than 100 µm, no mean value will
be displayed, and the measurement must be repeated.
Dr. Wilfried Bissmann
Carl
Zeiss
Ophthalmic
Instruments Division
Phone +49 3641 64 2058
Fax +49 3641 64
2917
E-mail bissmann@zeiss.de
The
telephone hotline +49 3641 64
2030
________________________________________________________________
Last
revision : July 24, 2000