Axial length measurement in pseudophakic patients -

2 peaks occur in some cases

The measurement of pseudophakic patients is problematic with ultrasonic biometers, if not indeed impossible in some cases.

Optical biometery allows the easy, problem-free performance of these measurements.

All you have to do is select the appropriate mode in the menu and conduct the measurement in the same way as for phakic eyes.

2 peaks can occur (see illustration) in strongly reflecting intraocular lenses (e.g. the Acrysof). The first is the a reflection from the IOL, while the second indicates the correct axial length. In these cases the measurements must be corrected manually.

Two peaks in the measurement of an Acrysof intraocular lens